The results of Hot Rod phantom with Xtrim PET system in PNP Co.

The results of Hot Rod phantom with Xtrim PET system in PNP Co.
Dedicated hot rod phantom was used for tomographic performance evaluation of the system

The main part of the phantom is a cylinder with 35 mm height and 32 mm diameter. Inside it, various fillable rods were drilled in 6 sections. The radius of rods ranges from 1.8 to 2.8 mm with steps of 2 mm from section to section. In each section, the distance between every 2 rods is equal to the radius of the rods in that section. The image of phantom without resolution recovery was optained

The results of Hot Rod phantom with Xtrim PET system in PNP Co.1

The results of Hot Rod phantom with Xtrim PET system in PNP Co.2

The results of Image Quality phantom with Xtrim PET system in PNP Co.

The results of Hot Rod phantom with Xtrim PET system in PNP Co.3

The results of Hot Rod phantom with Xtrim PET system in PNP Co.4

More News

25 April 2025
،he third GammaProbe device produced by the knowledge-based company Parto Negar Persia was officially delivered and put into operation at Khatam Al-Anbia Specialty Hospital.
18 April 2025
According to a resolution passed by the Cabinet of Ministers, the relative value of codes 100710 and 100715—which pertain to sentinel lymph node excision during mastectomy procedures—will increase by approximately 50K starting from the beginning of 1404.
05 March 2025
Parto Negar Persia Company has been honored with the award and certificate of appreciation as the outstanding sales representative in the eleven editions of the Iran-Made Exhibition.